Click here for the quiz.
25 multiple choice questions
Start at 6:00pm end by 7:00 pm
Wednesday, October 27, 2010
Sunday, October 3, 2010
Practice Questions (Due October 13th)
Choose the best response for each of the following statements.
For a sagital MR image of the lumbar vertebrae this is the gradient that is used for frequency encoding.
Procedure Slice Selection Phase Encoding Frequency Encoding
Sag. Knee
Coronal neck
Axial Lumbar
Sag. T. Spine
For a sagital MR image of the lumbar vertebrae this is the gradient that is used for frequency encoding.
- X
- Y
- Z
- X
- Y
- Z
This is controlled by the slope of the slice select gradient.
- Image matrix
- Slice Thickness
- Field of View
This is the precessional frequency of hydrogen when the field strength of the magnet is 10,000 Gauss.
- 21.28 MHz
- 42.57 MHz
- 63.86 MHz
- 10,000 MHz
This is one way of achieving a thin slice in MRI. (reduce slice thickness).
- use a steep slice select gradient slope
- use a broad transmit bandwidth.
- use a narrow receive bandwidth.
- Both A and C
This gradient is switched on during the collection of the MR signal (during the echo)
- slice select gradient
- phase encoding gradient
- frequency encoding gradient
Fill in the blanks with X, Y or Z based on the correct gradient.
October--Lecture Notes 1
Image Density and Contrast
Based on previous discussions, we know that different tissues in the body will interact with radiofrequency (RF) energy in different ways. For example, fat will give up the RF energy it received during the 90 degree RF pulse very quickly (short T1 and an even shorter T2 time). Conversely, water takes longer to give up the absorbed RF energy (relax), and therefore has longer T1 and T2 times.
The T1 curves shown above compare the recovery times for fat, white matter, grey matter and CSF. The horizontal axis on the graph represents repetition time (TR) in seconds. Signal intensity can be measured (read by the machine) at any time.
Based on the TI curves in the graph shown above, answer the following questions:-
If signal intensity is measured at the time labeled T1 CSF:
- What is the contrast between grey and white matter on the resulting image?
- Which two tissues will have the highest contrast?
The density of each tissue varies based on when the signal intensity is measured. For the TE shown in the diagram below (slightly less than 0.2 seconds or 200ms), CSF will appear darker than all the other tissues because it has the lowest signal intensity.
Spatial Localization in MRI
We know how a MR signal is generated and how appearance of different tissues on the image is determined (image weighting), but how does the machine determine where the signal is coming from (spatial localization) ?
The machine needs this spatial information in order to form the image. Each voxel in the patient provides information for the corresponding pixel in the image.
The anatomy is broken down into slices of a particular thickness, and then further divided into voxels based on the scan matrix. |
To localize the voxels, spatial information needs to be encoded into the MR signal.
Three steps are required :
- First of all, the desired slice must be selected
- Then, spatial information is encoded along the rows
- Finally, spatial information is encoded along the columns